Hashir Aazh and Brian C. J. Moore

Incidence of Discomfort During Pure-Tone Audiometry and Measurement of Uncomfortable Loudness Levels Among People Seeking Help for Tinnitus and/or Hyperacusis

Purpose

The aim of this study was to assess the proportion of patients seen in a tinnitus and hyperacusis therapy clinic for whom presentation levels based on the British Society of Audiology (BSA)–recommended procedures for pure-tone audiometry and determination of uncomfortable loudness levels (ULLs) exceed ULLs, leading to discomfort during administration of these procedures.

Method

This was a retrospective cross-sectional study of 362 consecutive patients who attended a National Health Service audiology clinic for tinnitus and/or hyperacusis rehabilitation.

Results

For 21% of the patients, presentation levels based on the BSA procedure for pure-tone audiometry exceeded the ULL for at least 1 of the measured frequencies (excluding the first frequency tested, 1 kHz): 0.25, 0.5, 2, 3, 4, 6, and 8 kHz. For 24% of patients, the starting presentation level of 60 dB hearing level recommended for determination of ULLs exceeded the ULL for at least 1 frequency.

Conclusion

The starting presentation levels used for pure-tone audiometry and measurement of ULLs should be lower than those recommended by the BSA for people with tinnitus and hyperacusis.


Research source
For Hearing Healthcare Professionals